CVS4200 / BVS4300 Stationary Samplers

3. Press an ARROW key. Continue until
Composite is shown on the display (for
storage in single container).

4. Press the ENTER key.

The display will respond with the message
“Terminate After 0003000 Samples”. The
numerical value will be whatever value was
last placed in the controller’s memory,
usually after previous programming. To
keep the previous value, press ENTER, or,
to set a new value, use the ARROW keys to
advance the flashing prompt to the desired
location and replace the digits under the
prompt by using the digits (0-9) on the
Touchpad. Each new entry will
automatically advance the prompt to the next
location. In this way, the entire 7 digit
number can be changed. The ARROW keys
can be used to skip already correct digits, in
either direction.

When the 9 digit number is correctly
entered, press ENTER.

The display will echo the last entry with
PROGRAM TYPE <ENTERED>.

The controller has now been given a new
value for PROGRAM TYPE. The new
value resides in the NEW ENTRIES area of
the controller memory. To make these
changes active, press the RESTART key;
press it again to confirm your choice. The
controller will then be set to perform as a
Composite Sampler in conjunction with the
parameters programmed under the START
DELAY
and SAMPLE INITIATION
variables.


Summary of Sequence:
SET, PROGRAM TYPE, ARROW(S), ENTER, #######, ENTER,
RESTART, RESTART.