CVS4200 / BVS4300 Stationary Samplers
value is required, the value has been 
changed or the value is zero, pressing 
ENTER will advance the day of the week. 
Therefore... 
Press ENTER. Repeat the above procedures 
until all the required containers, on their 
respective days, have been allocated, or the 
7 days of the week are all selected. The total 
number available is never allowed to be 
more than the amount registered as the two 
digit representation of the hardware 
configuration in the “Total bottles 00” entry. 
The display will echo the last entry with 
PROGRAM TYPE <ENTERED>. 
The controller has now been given a new 
value for PROGRAM TYPE. The new 
value resides in the NEW ENTRIES area of 
the controller memory. To make these 
changes active, press the RESTART key; 
press it again to confirm your choice. The 
controller will then be set to perform as a 
Daily Cycle Sampler in conjunction with the 
parameters programmed under the START 
DELAY and SAMPLE INITIATION 
variables. 
 
Summary of Sequence: 
SET, PROGRAM TYPE, ARROW(S), ENTER, ARROW(S), ENTER, #, 
ENTER, #, ENTER, #, ENTER, #, ENTER, #, ENTER, #, ENTER, #, 
ENTER, RESTART, RESTART. 
7.6.6.4 PROGRAM TYPE - Daily Cycle for Dual Station
Dual Station – Flip Flop Application Only (Single Controller, Two Metering 
Chambers) 
The following sequence of entries are made 
on the Touchpad to describe how the 
Sampler controller is to store the samples it 
takes, when the hardware specified is 
configured to deliver Samples from two 
separate sources. The ACTIVE SETTINGS 
are not being altered. 
55