5. Transport, packaging and storage

5.1 Transport

Check instrument for any damage that may have been caused by transport. Obvious damage

must be reported immediately.

5.2 Packaging

Do not remove packaging until just before mounting.

Keep the packaging as it will provide optimum protection during transport (e.g. change in

installation site, sending for repair).

5.3 Storage
Permissible conditions at the place of storage:

Storage temperature:

Instruments without built-in transmitter: -40 ... +85 °C

Instruments with built-in transmitter:

see operating instructions of the corresponding

transmitter

Humidity: 35 ... 85 % relative humidity (no condensation)

Avoid exposure to the following factors:

Direct sunlight or proximity to hot objects

Mechanical vibration, mechanical shock (putting it down hard)

Soot, vapour, dust and corrosive gases

Potentially explosive environments, flammable atmospheres

Store the instrument in its original packaging in a location that fulfills the conditions listed above. If

the original packaging is not available, pack and store the instrument as described below:

1. Wrap the instrument in an antistatic plastic film.

2. Place the instrument, along with shock-absorbent material, in the packaging.

3. If stored for a prolonged period of time (more than 30 days), place a bag, containing a

desiccant, inside the packaging.

WARNING!

Before storing the instrument (following operation), remove any residual media. This

is of particular importance if the medium is hazardous to health, e.g. caustic, toxic,

carcinogenic, radioactive, etc.

6. Commissioning, operation

WARNING!

When the thermometer is mounted, the temperature must not drop below the allowed

operating temperature (environment, medium) or exceed it, even when taking

convection and heat radiation into account!

WARNING!

Thermometers must be earthed if dangerous voltages could be expected at the

connection wires (caused, for example, by mechanical damage, electrostatic

discharge or induction)!