Commissioning, operation – WIKA TR10-J User Manual

Page 14

Advertising
background image

11208287.06 10/2013 GB/D/F/E

14

WIKA operating instructions resistance thermometers, thermocouples

GB

Thermocouples with cable connection or male connector

6. Commissioning, operation

Lemosa connector,

(male) on cable

Binder connector,

(male) on cable

(threaded plug connection)

Single

thermocouple

Colour coding of

the wire ends see

table

Cable

Dual

thermocouple

Colour code of cable

Positive and negative terminal are marked. Two

miniature size thermo connectors are used with

dual thermocouples.

Thermo connector

Type of sensor Standard

Positive terminal

Negative terminal

K

DIN EN 60584

green

white

J

DIN EN 60584

black

white

E

DIN EN 60584

violet

white

T

DIN EN 60584

brown

white

N

DIN EN 60584

pink

white

Advertising