3. Specifications

3.1 Resistance thermometer
Sensor connection method

2-wire

3-wire

4-wire

Sensor tolerance value per DIN EN 60751

Class B

Class A

Class AA

The combinations of a 2-wire connection with class A or class AA are not permissible, since the

lead resistance of the measuring insert negates the higher sensor accuracy.

Basic values and tolerance values

Basic values and tolerance values for the platinum measurement resistances are laid down in

DIN EN 60751.

The nominal value of Pt100 sensors is 100 Ω at 0 °C. The temperature coefficient α can be stated

simply to be between 0 °C and 100 °C with:

α = 3.85 ∙ 10

-3

°C

-1

The relationship between temperature and electrical resistance is described by polynomials,

which are also defined in DIN EN 60751. Moreover, this standard specifies the basic values in

°C steps in tabular form. Moreover, this standard specifies the basic values in °C steps in tabular

form.

Class

Temperature range

Tolerance value in °C

Wire-wound (W)

Thin-film (F)

B

-196 … +600 °C

-50 … +500 °C

±(0.30 + 0.0050 | t |)

1)

A

-100 … +450 °C

-30 … +300 °C

±(0.15 + 0.0020 | t |)

1)

AA

-50 … +250 °C

0 … 150 °C

±(0.10 + 0.0017 | t |)

1)

1) | t | is the value of the temperature in °C without consideration of the sign.

Bold: Standard version

For further technical information, see WIKA data sheet and the technical information sheet

IN 00.17 “Usage limitations and accuracies of platinum resistance thermometers per

EN 60751: 2008”.