BNC 6040 User Manual

Page 15

Advertising
background image

15




SECTION 2

OPERATING INFORMATION


General

POWER UP

When power is first applied, the mainframe performs a test to determine if a module is installed. If a
module is present, the limits for each module dependent parameter are transferred to the mainframe.
The memory is checked for any errors and the parameters in memory 0 are loaded into the mainframe.

If a module is not present, the parameters are set as indicated in Table 2-1.

Table 1-4. 6040 Mainframe Default Settings





















LCD Power On Sequence

When the 6040 is first turned on, the display will show the software version number. If a module is
present, the LCD will then momentarily display "MEMORY CHECK," followed by the module I.D.
display. Without a module, the display will read "6040 No Plug-In."





MODE:

Pulse


TRIG:

Single Cycle (with other values set as follows)

Internal Trigger Rate = 1 kHz

External Trigger Threshold = 0 V

External Trigger Slope: +


TIMING:

Width = 1 µs

Delay = 1 µs

Single Pulse


GPIB/RS-232: IEEE-488 Address = 6

Baud Rate = 1200

Full Duplex

Remote Enabled