 :DUUDQW\ FODLPV DQG OLDELOLW\

Please read these Operating Instructions carefully and in their entirety. IKA

will only

consider itself responsible for the safety, reliability and performance of the device if

the device has been operated in accordance with the Operating Instructions

only persons authorized by the manufacturer have performed repair and main-

tenance work on the device

only original parts and original accessories have been used in repair work.

The calorimeter system may only be opened by a customer service center.

If service is required, we recommend that you contact our customer service depart-

ment.

Furthermore, we refer you to the applicable safety conditions and accident require-

ments.

IKA

is not responsible for damages or costs resulting from accident, improper use

of the device or impermissible modifications, repairs or renovations.

 6\VWHP SURSHUWLHV

The C 2000 calorimeter system is routinely used for determination of gross calorific

values of solid and liquid substances. The accessories of the system ensure cus-

tomized adaptation to laboratory tasks (see Chapter 13).

The system is characterized by the following features:

Relieves you of routine tasks through automated measurement procedure

Integrated oxygen filling

Automatic detection of decomposition vessel (DV)

Operation without the cooling unit: Connection to a water faucet; temperature

range 12°C - 28°C; water consumption per measurement about 4 l; water pres-

sure 1 bar to max. 1.5 bar; at higher pressures, use C 25.

Operation with active cooling unit (for example IKA

KV 500, optional)

Measurement and calculation of gross calorific value according to DIN 51900,

ISO 1928, ASTM D240, ASTM D4809, ASTM D5865, ASTM D1989, ASTM

D5468, ASTM E711

Calculation of net calorific value according to DIN 51900, ASTM D240,

ASTM D4809, ASTM D5865, ASTM D1989, ASTM D5468, ASTM E711

Measurement range: max. 40,000 J

(This corresponds to an increase in temperature in the inner vessel of about

5 K)

SDUWV FRQGXFWLQJ

HOHFWULFDO SRZHU