IKA C 2000 control User Manual

Page 53

Advertising
background image

IKA

-WERKE C 2000 basic / control

Ver. 04 08.07

3DJH 

The message "Bomb securely closed?" will appear. Ensure that the decomposition

vessel is properly closed and confirm with OK.

The ignition wire in the decomposition vessel completes the electrical circuit. The

calorimeter is ready to start. The message

%RPE

changes to a display of the func-

tion key assignment

6WDUW

If the

6WDUW

function key does not appear, please check

the ignition wire of the decomposition vessel.

d

Push the

6WDUW

button.

Each time 1000 ignitions have been performed using a given decomposition vessel,

the following message will appear:

1000 ignitions performed with Bomb x

Inspect decomposition vessel

or contact IKA service

This indicates that the decomposition vessel has reached a maintenance point and

that a safety check must be carried out. Confirm this message by pressing

  7$%

and

2.

in sequence.

This message does not release the user from the responsibility of also continuously

checking the decomposition vessel for wear and carrying out safety inspections as

required.

In

([SORVLYH

mode, please observe in addition the following:

As many as 20 decomposition vessels per device can be used. As many measure-

ments can be prepared as there are decomposition vessels available.

1

cap screw

2

filling head

3

ignition contact

4

spring element

VXVSHQGLQJ WKH

GHFRPSRVLWLRQ

YHVVHO LQWR WKH

ILOOHU KHDG RI WKH

PHDVXULQJ FHOO

FRYHU









Advertising