The following apply to evaluating successful calibrations:

05)> @

The average relative error < 0.2 % (according to ISO 1928)

'LII

The deviation range as a percentage < 0.4 % (according to

DIN 51900)

Other criteria may also apply depending on the standard you are using. The values

given above, however, will be sufficient to fulfill customary requirements of accuracy

for a calorimeter.

k

Use

7DE

to place the cursor on the

>±!@

button and confirm with

2.

or press



. The

average value of the selected calibration experiments will thus be assigned to the

calorimeter system as the system heat capacity or C value. If you place the cursor

on

>±@

press

2.

, you can enter the system heat capacity manually in the

& YDOXH

field.

l

Place the cursor on the experiments that were not used to calculate the average

value and delete them with

'HO

cc

Exit the dialog box with

2.

The system calibration is thus complete and you can

now continue with the determination of gross calorific values.