IKA C 2000 control User Manual

Page 81

Advertising
background image

IKA

-WERKE C 2000 basic / control

Ver. 04 08.07

3DJH 

0HVVDJH

&DXVH

5HPHG\

3UHOLPLQDU\ H[SHULPHQW

!  PLQ

or

PDLQ H[SHUL

PHQW !  PLQ

The combustion experiment is

lasting too long. Isoperibolic

measurements are stopped after

13 minutes in the “Stabilize”

phase or after 16 minutes in the

main experiment.

Check:

the operation of the stirrer drive

(Maintenance menu:

6WLUUHU

RQRII

).

temperature equalization of the

calorimeter to the working tem-

perature with the Maintenance

menu:

7HPSHUDWXUH HTXDOL]DWLRQ

The seal of the decomposition

vessel (see C 5010/C 5012 Op-

erating Instructions).

!!

7HPSHUDWXUH [[[[ ƒ&

Malfunction while recording the

temperature

The temperature display does

not change any more.

Open the information window and

look at the temperature displays

7HPS

and

7HPS

. If Temp1 or

Temp2 do not change within 60 sec-

onds, there is an error in tempera-

ture recording.

Both values are constant: In the

Maintenance menu, execute the

7HPS ,QLW

command.

If you cannot reach the Maintenance

menu or if you do but this does not

provide any help, turn the calorime-

ter off and back on again.

Only Temp2 is constant: In the

Maintenance menu, execute the

&RROLQJ ZDWHU

command. After 60

seconds at the most, the displayed

cooling water temperature should no

longer be constant. If you cannot

reach the Maintenance menu or if

you do but this does not provide any

help, use another method to check

the cooling water temperature. The

cooling water temperature must fall

within the range of 5.xx°C to 28°C

before it is possible to continue

working.