When you turn on the calorimeter, the splash screen first appears (the cover of the

measuring cell opens automatically and the stirrer starts up for a few seconds).

The current assignment of the function keys appears in the footer, and the current

system message appears in the header.

Make certain that the water connection for the device has been opened and if there

is an additional cooler unit that it is turned on.

After you confirm the splash screen (or if you do not then automatically after 10 sec-

onds), the system check is started. The system check verifies whether the cooling

water is flowing through the calorimeter (at least 30 seconds).

Next the system tests whether the measured cooling water temperature is suitable

for stable operation in the selected operating mode (at least 150 seconds). During

this time, the measured cooling water temperature that is suitable for the mode is

displayed in the left selection table, while the right selection table designates the last

operating mode that was used.

VSODVK VFUHHQ

YHULILFDWLRQ RI

IORZWKURXJK