Controller display messages – C.B.S. Scientific DGGEK-2401 User Manual

Page 12

Advertising
background image

12

Cipher DGGE Instructions 2/28/14

2 3 4 h Local Lockout Feature

This feature enables the user to lock all controls on the controller. While the feature is acti-
vated, the unit will remain running at the current settings. To activate the local lockout feature,
press and hold the Select/Set Knob for 10 seconds. Once locked, the screen will read LLo.
When locked, the setpoint decimal point will not flash as usual. Press and hold the Select/Set
Knob again for 10 seconds to unlock the controls. Once unlocked, the screen will read CAn.

2 3 4 i Stick-on Strips for Preset Buttons

The Standard Controller is supplied with two
stick-on strips that can be applied beneath the
Preset Buttons. These strips are removable and
can be reapplied. The user can write the set
point temperature associated with each Preset
Button on this strip. The use of an erasable
medium, such as dry erase marker or flair tip
pen is recommended. The former can be wiped
off with a dry cloth or tissue; the later can usu-
ally be removed with simple glass cleaner and a
cloth or tissue. Do not use a permanent marker
or ballpoint pen. Always do an ink test before
writing on these strips.

2 3 5

Controller Display Messages

The stick-on strip fits beneath the controller’s
Preset Buttons. It is removable and reapplicable.





Display

Description

Action Required

….

Standby mode

Normal — Indicates that the Circuit Breaker/Power Switch is ON
and the Controller Power Switch is OFF

tx.xx

Power up self-test

Normal — Appears momentarily at startup

oCx.x

Calibration offset value

Normal — Current calibration offset value; refer to section 6.1

Hxxx

Software High Limit value

Normal — Current Software High Limit value; refer to section 5.6

Axxx

Auto-Refrigeration set point value

Normal — Current Auto-Refrigeration set point; refer to section
5.10 (Appears only on Refrigerating/Heating Circulators)

E-H1

Software High Limit set point too low

Error — The value entered is below the control temperature set
point. Refer to Setting Software High Limit, section 5.6

FLt 1

Software High Limit value exceeded

Error — Set a Software High Limit value higher, then turn main
power to the unit OFF and back ON; refer to section 5.6

FLt 2

EEPROM reset

Error — Turn Circuit Breaker/Power Switch OFF, hold P3 Button,
and then turn the Circuit Breaker/Power Switch back ON

FLt 3

Safety Set temperature exceeded

Error — Check fluid level
Check fluid temperature and set point
Ensure that OTP set point is higher than fluid set point or
increase to maximum
Turn Circuit Breaker/Power Switch OFF, press Safety Set Reset
Button, and then turn the Circuit Breaker/Power Switch back ON

FLt 4

Heating Triac failure

Error — Service required

FLt 5

Probe failure

Error — Service required

FLt 6

i2c error — Communication failure to
modulation board

Error — Service required (Appears only on Refrigerating/Heating
Circulators)







Calibration allows the user to match the Controller’s bath temperature display to an external
reference thermometer. Calibration is performed as follows:

Set the desired operating fluid temperature set point and allow temperature to stabilize.

Press the P2 and P3 simultaneously and release and repeat until the display reads (oCx.x). Press
P1 and hold until (Cal) is displayed. This will take about 2 seconds.

At one second intervals, the displayed value will alternate between the actual bath fluid temperature

and the current offset value, which is the difference between the factory calibration setting and the
user’s reference temperature sensor. The maximum offset is ±0.9°C from factory calibration.
To change the calibration offset value, rotate the Select/Set Knob until the display matches the

reading on the reference temperature sensor. The display will continue to alternate between the
offset value and the calibrated display temperature.

Press the Select/Set Knob or the P1 Button to accept the entered value. When the new calibration is
stored and the mode is exited, (dONE) will appear on the display.

NOTE: The displayed offset value will also be accepted if there no activity for 20 seconds.




The heater should be kept clean. If deposits build up on the heater, they may be removed by
scrubbing with a non-metallic (plastic) abrasive pad. Do not use steel wool.