Cipher DGGE Instructions 2/28/14

DGGEK Unit Set-Up, con’t

2 4 Setting Buffer Temperature

Warning: Do NOT turn on Heater/Stirrer until tank has been filled with buffer!

1. Turn Heater/Stirrer “On” and set temperature to 60°C according to

Section

2 3 4 f Setting the Set Point Temperature (page 11)

2 5 Connecting the Power Supply to Safety Interlock

2. Position EPS-300 X Power Supply on a shelf above the tank. Refer

to Figure 2-2 to complete the steps below:

a. Attach the power supply to the safety interlock.

b. Attach the safety interlock to a wall (mains) outlet.

c Do NOT attach power leads (from power supply to safety

interlock) until gels are loaded and installed in the tank