C.B.S. Scientific DGGEK-2401 User Manual

Page 32

Advertising
background image

32

Cipher DGGE Instructions 2/28/14

SECTION 10

Maintenance of Equipment

10 1 Care and Handling

The plastic components of the DGGE Systems are fabricated from acrylic

and polycarbonate. Electrodes and connectors are made from pure platinum,

stainless steel, and chrome plated brass. As with any laboratory instrument,

adequate care ensures consistent and reliable performance.

After each use, rinse gel cassettes, spacers and combs with de-ionized

water. Wipe dry with a soft cloth or paper towel, or allow to air dry. Whenever

necessary, all components may be washed gently with water and a non-

abrasive detergent, and rinsed and dried as above. Never use abrasive

cleaners, glass cleaning sprays or scouring pads to clean the components, as

these will damage the unit and components.

Additional precautions:

• Do not autoclave or dry-heat sterilize the apparatus or

components.

• Do not expose the apparatus or components to phenol,

acetone, benzene, halogenated hydrocarbon solvents or

undiluted alcohol's.

• Avoid prolonged exposure of the apparatus or components to

UV light.

• Do NOT treat with diethylpyrocarbonate (DEPC)-treated water

for extended periods at 37°C. A brief rinse with DEPC-water

is sufficient after a thorough wash.

10 2 Maintenance

The following inspection and maintenance procedures will help maintain the

safety and reliable performance of the Cipher DGGEK systems. Replacement

parts can be ordered by calling 858-755-4959 or by contacting your local

distributor.

• Banana plugs and power cords should be inspected regularly.

If the banana plugs become loose or do not feel friction tight

replace the plugs or power cords.

• Should power cord assemblies (connectors, wire or shrouds)

show any signs of wear or damage (e.g. cracks, nicks,

abrasions, or melted insulation), replace them immediately.

• The platinum wire is secured to the banana jack by compression

between a stainless washer and the jack nut. The nut/washer

interface should be tight and free of corrosion.

Advertising
This manual is related to the following products: