Cipher DGGE Instructions 2/28/14

SECTION 3

Gel Casting Options

3 1 Preparation/Cleaning of Glass Plate for Gel Casting

Hand wash both plates with a high quality lab detergent followed by a complete

rinsing with dH

2

O. Air dry or use a lint-free tissue. Spray/wipe the chosen

inner surfaces of the plate set with 95% ethanol and dry with lint-free tissue.

3 2 Gel Casting Techniques

A. Gel Wrap™ Gasket Casting Method.

B. Vertical gradient gel casting using GM-40 gradient maker and gravity flow.

C. Vertical gradient gel casting using GM-40 gradient maker and a Mini-pump.

D. Vertical gradient gel casting using GM-40 gradient maker, Mini-pump and

Multi-gel Caster.

NOTE: Prior to casting, mark the notched glass plate to designate which side is the

‘inside’ and ‘outside’. The same side of the notched plate should always sit against the

reservoir gasket.

A Gel Casting using Gel WrapGasket Casting method

For Vertical DGGE, use the set of spacers which do not have the small hole or channel

milled into the lower end, these are for casting

perpendicular gels.

1. Start by holding the rectangular back

plate with the rounded bottom corners

and start applying the gasket around one

side of the glass plate. Note: one side

of the “U” shaped gasket is flat, and the

other side has tubing that will act as a

seal around the spacers.

2. When applying the gasket over the

rounded corners of the back glass plate,

make sure the notches on the gasket

align with the rounded corners of the

glass plate. Once the gasket is pushed

over the bottom edge and corners, work

it down the remaining side.