Dggek-2001, Figure 2-1, 2 2 filling dgge tank with buffer – C.B.S. Scientific DGGEK-2401 User Manual

Page 7

Advertising
background image

7

www.cbsscientific.com

Figure 2-1

Heater

Stirrer

Buffer Siphon

Pump Assembly

Vapor

Shield

Thermometer

DGGE

Tank

Single

Cassetes

Comb (2)

Spacers

(2 pairs)

Glass Plate

Assembly(2)

Clamps

Mini-Pump

(Optional)

GM-40

Gelwrap

® (2)

 

Buffer recycling
manifold

2. Open the lid and confirm the following:

a) Two lengths of small bore silicone tubing for buffer cycling are already attached

to the buffer recycling manifold outlets

b) A single length of large bore tubing from the By-Pass Pump is connected to the

barb-fitting on the plastic manifold.

DGGEK-2001

2 2 Filling DGGE tank with Buffer

1) Fill the tank with running buffer (see Section 6 for buffer formulations) to optimal

level for number and type of cassettes used. Refer to chart below for exact buffer

volumes required: This requires 20 to 21.2 liters of buffer. See chart below:

Use distilled or RO water when making buffer
NOTE:
Buffer Displacement Tank Instructions:

With DGGE-1001 - fill buffer displacement tank, BDT-1000 with water and place in

the back of DGGE tank.

With DGGE-2001 or DGGE-2401- If using only 1 cassette, BDT-1000 can also be

used to save buffer but must be ordered separately.

Model Cat. #

Cipher DGGEK-1001

Cipher DGGEK-2001

Cipher DGGEK-2401

Optimal Buffer

Volume

21 Liters -1 dual cassette

21.5 Liters - 1 single cassette

21 Liters - 2 single cassettes

21 Liters - 1 dual cassette

20 Liters - 2 dual cassettes