Models 7401 and 7403 do not have sliding shelves, but a sliding shelf accessory can be
purchased.


Leveling the Unit

Place a level on the middle shelf of the incubator. Adjust the feet until the unit sits level
left to right and front to back. Even if the unit is level without adjustment, the leveling
feet should still be lowered to avoid the cabinet moving while opening and closing the
outer door & prevent a flat spot from forming on the casters.


Connecting the Drain






The incubator can have humidity either by an evaporative pan or injecting water. To
remove excess water that is not humidified or water that can be spilled when filling the
evaporative pan, connect the drain fitting and tubing supplied with the incubator to a
local floor drain, minimizing the risk of contamination and corrosion.








If a local floor drain is not available, a variety of accessories are available through
CARON customer service. These accessories can also be viewed at

www.caronproducts.com


Models 7401-25/33, 7402-25/33 and 7403-25/33 are non-humidified chambers and have
a 1/2” plug installed into the drain connection. The drain will be required when a
humidity option is purchased with the chamber.




When using a pressurized water source, failure
to connect the unit to a drain could result in
facility flooding.

The drain line relies on gravity to remove water
from the chamber. The drain line must remain
below the chamber to drain properly. Kinks or
elevations in the drain line above the cabinet
drain will not allow the chamber to drain.