7400 Series Operations Manual

Rev A 07/25/14

Page 74 of 83

SECTION 12 – PREVENTATIVE MAINTENANCE


Your CARON incubator has been robustly designed to minimize performance problems.
However, regular maintenance is very important for continuous trouble free operation.

As a general rule, CARON recommends an annual calibration check of the temperature,
humidity, CO2 systems. CARON offers a full range of on-site calibration and validation
services. We also offer preventative

maintenance contracts on our equipment. Contact

our service department for details at 740

-373-6809 or visit us on the web at

www.caronproducts.com

.


Recommended Daily Maintenance Checks

 Check the Temperature, humidity, and CO2 displays versus set-points.

 Check for and correct any alarm condition.


Recommended Monthly Maintenance Checks

 Check to ensure the drain in the bottom of the unit is draining properly.

 Check front air intake filter. If it is dirty replace it with CARON part number

FLTR301. Washing the filter will result in poor performance.


Recommended Annual Maintenance Checks

 Disinfect all interior surfaces with a general purpose laboratory cleaning agent.

 Perform a complete calibration of the temperature, humidity, and CO2 systems.

 A full validation is recommended for GMP facilities each time a unit is installed,

moved or undergoes significant repair. Contact CARON’s service department to
schedule on-site validation.