7400 Series Operations Manual

Rev A 07/25/14

Page 33 of 83

Do not immerse the sensor in any type of
cleaner as damage to the sensor may occur.

6) Insert the orange cable into the grommet where the sensor was removed so that it

is easily accessed when the sensor is replaced following the decon cycle.

7) Replace the CO2 sensor plate and turn the incubator power switch back on.
8) The infrared CO2 sensor can be disinfected using isopropanol. Spray the cleaner

on a soft clean cloth and wipe the sensor.




Starting the Decon Cycle process, from the home screen,

Press the

(Settings) button.

Press the

(Decon Cycle) button on the left side of the screen.

Settings
button

Decon Cycle