Caron 7403-33 User Manual

Page 47

Advertising
background image

7400 Series Operations Manual

Rev A 07/25/14

Page 47 of 83

SECTION 7 – CALIBRATION


The temperature and humidity systems can all be calibrated as necessary. CARON
recommends an annual calibration check of each system. Before making a calibration
adjustment, allow the cabinet to stabilize a minimum of 12 hours from a power off
condition. If the unit has been in operation, allow a minimum of 3 hours of stable
operation at all set-points. If you do not have the appropriate reference instruments to
perform calibration, contact CARON’s service department for on-site calibration at

[email protected]

. Caron also provides validation services which ensures

that the unit is functioning properly according to IQ, OQ and PQ protocols which satisfy
FDA guidelines for qualification verification of equipment.





The Calibration Screen

To get to the calibration screen from the home page:

Main screen with HUMD304, HUMD307 option

Press the

(Settings) button.









Be sure that all reference instruments are
calibrated to an appropriate standard.

Settings
Button