Connect each incubator to a grounded circuit.
Failure to do so could result in electrical shock.
If the unit is equipped with a built in gas guard system GASG302, there will be 2 gas
inlets. Each of the inlets should be connected to an individual gas tank as described
above.
Connecting Electrical Power
The unit requires a dedicated electrical outlet. See table below for model specific power
required and connection.
Model #
Power Requirements
Plug Connection
-1
115V, 60Hz, 16A FLA
NEMA 5-20
-2
230V, 60Hz, 10A FLA
NEMA 6-15
-3
230V, 50Hz, 8A FLA
CEE 7/7
When the required electrical connection is available, plug the provided power cord into
the unit and the electrical outlet.
Cell Roller Ready (Models 7402 & 7403)
These models have a cell roller, a reinforced floor, cell roller ramp, and interior outlet.
These models were designed to hold a 3, 5 or 7 tier Wheaton R2P cell roller.
Support braces have tracks to guide the cell roller wheels into the incubator. The cell
roller ramp has small tabs at the end that fit into an opening at the end of each cell roller
support brace.
The ramp is designed to support the weight of the cell roller as it is pushed into the
incubator. Once the cell roller is pushed into the incubator, remove the ramp and save it
in a convenient place should it need to be reused to remove the cell roller.
Cell Rollers can be heavy and awkward to
handle. Ensure that sufficient resources are
available to safely move the product.