Caron 7403-33 User Manual
Page 21

7400 Series Operations Manual
Rev A 07/25/14
Page 21 of 83
Installing Drain Water Pump (PUMP301)
In applications where a floor drain is not available and a CARON water recycling system
CRSY102 is not being used, a drain pump can be purchased to pump any excess
condensate from the chamber to a local sink or drain. The pump is located near the
middle of the back of the chamber. Connect the supplied tubing from the pump to the
sink / drain. The tubing may be run vertically into a ceiling but should not exceed 15 feet
height. The pump is equipped with a small reservoir on the bottom of the pump with an
internal level switch that will automatically turn the pump ON when it is full to drain the
water out of the reservoir and into a floor or sink drain.
Pump Outlet
to Sink or Floor
Drain
Pump Inlet
from Chamber
Drain
Reservoir with
Internal Level
Switch