Caron 7403-33 User Manual
Page 38

7400 Series Operations Manual
Rev A 07/25/14
Page 38 of 83
The factory default “master tank” is Tank 1. When the appropriate gas pressure is
supplied to both tanks, the master tank will always be used as the gas source. The unit
will swap from the master tank to the alternative tank whenever a low gas pressure
condition is detected.
Ultraviolet Germicidal Lamp (LGHT602)
Replacing UV Light (optional accessory)
1. Turn off chamber and unplug power cord.
2. Remove the incubator top access panel. The housing is located in the back of the
incubator.
3. Unclip green wire with ground clip from UV light housing.
4. Pull UV light housing cap from UV light housing. Connected UV lamp will come
out with it.
Before removing access panel(s), disconnect
electrical power.
Avoid exposure to direct or reflected germicidal
ultraviolet rays. Germicidal ultraviolet rays are
harmful to the eyes and skin.
UV Light Housing Cap
Green Wire with
Ground Clip