Caron 7403-33 User Manual

Page 19

Advertising
background image

7400 Series Operations Manual

Rev A 07/25/14

Page 19 of 83

Use only distilled or deionized water with a
resistivity between 50K

-CM and 1M-CM and

a pH of greater than 6.5. Using water outside
this range will void your warranty.

Do not use water that contains chloramines.
Chloramines can damage internal rubber
gaskets resulting in leaks.

Connecting to Humidity System (HUMD304, HUMD307)















If these water sources are not available contact CARON customer service.

A water inlet fitting on the back of the unit and ¼” black tubing are provided to connect
the water supply to the incubator. Connect an appropriate water supply to the fitting.
Incoming line pressure should be regulated to not exceed 80 psi.

CRSY102 or BOTL301 are accessories that can be used as a water supplies. If a
Condensate Recirculator CRSY102 water recycling system was purchased as a water
supply, refer to its user’s manual for proper installation of the water supply.