Replacement instructions, Core sub-assembly removal and replacement, Handle / piston removal and replacement – FasTest MediMate Connector Body User Manual

Page 2

Advertising
background image

REPLACEMENT INSTRUCTIONS

Roseville, MN

Ph: 1-800-444-2373
Fax: 651-645-7390
www.fastestinc.com

WP066 10/6/2010

®

HANDLE

CLEVIS PIN

RETAINING RING

ACTIVATION PISTON

AND SPRING

BODY

SHAFT THREADS

CORE SUB

ASSEMBLY

Core Sub-Assembly Removal and Replacement:












1. To remove the core sub-assembly from the old body, place the body of the MediMate

connector in a clamp or vise.


2. Utilizing flats on shaft, unscrew shaft from the body. Use care so not to scratch or damage

the connector components. The core sub-assembly should easily slide out of the body. If
the components do not remove easily, gently remove parts by hand. Do not use any sharp
instruments.










3. Make sure the core sub-assembly is dry and clean. If necessary, blow-off the assembly

with clean dry air.

4. Apply a drop of LocTite 242 to the shaft threads. Use care to assure that the LocTite does

not contaminate any other components.









5. Place the core sub-assembly into the new body and tighten down with wrench until snug.

Do not over tighten.

Handle / Piston Removal and Replacement:

1. Remove the retaining ring from the clevis pin.
















2. Slide the clevis pin out through the handle and body.

















3. The activation piston will now be free and easily removed from the old body. Remove the

piston and the piston spring and transfer to the new body.


4. Orientate the handle into the new body and slide the clevis pin through the body and

handle.

5. Replace the retaining ring into the clevis pin.

SHAFT FLATS

SHAFT

SHAFT THREADS

RETAINING RING

CLEVIS PIN

ACTIVATION PISTON
AND SPRING

HANDLE