Rotavac valve tec – Heidolph VAC control automatic User Manual

Page 43

Advertising
background image

43

E

1.3. Vacuum supply

-

Connect vacuum with stud (12)


1.4. Electric branching

-

Before connecting with your local power supply make sure that:

mains voltage is identical with data plate (refer to "safety Information"; refer to "General" if
plug and outlet do not fit together.

-

Connect power cord with outlet.


2. ROTAVAC valve tec

Caution: ROTAVAC vario tec may be used together with the LABOROTA 4002/4003
evaporator only !


2.1. Set-up

-

Locate items on solid and plain surface. Item must be horizontal.

2.2. Install cooler holder and condensate cooler

-

If your ROTAVAC valve tes is used together with the
ROTACOOL reflux cooler, you may directly place your
ROTAVAC valve tec on top of the ROTACOOL.

-

Remove fairing (17) by unthreading two screws

-

Install bracket (1) for condensate cooler (2) on the rear of the
vacuum pump in a slot provided to this end. Introduce 2 ea.
square nuts in slot.

-

Re-install fairing (17).

-

Remove coupling nut (10) from
condensate cooler (Attention,
retainer underneath coupling nut!)

-

Locate condensate cooler (2) in
holder

-

Install receiving flask (4) and secure
with clamp (5).

-

Fit red coupling ring (10) and seal (9)
on right-hand end of PTFE hose (7).
Plug this line assembly in glass
thread (18) of aftercooler (2). Plug
left-hand over outlet stud (6), pull
coupling nut tight.

-

Connect cooling water with fittings
(11), exhaust air with fitting (14), as
required.






14

2

1

6

17

4

5

10

7

18

11

12

ROTACOOL

Advertising
This manual is related to the following products: