Specifications – Heidolph VAC control automatic User Manual

Page 66

Advertising
background image

66

E

SPECIFICATIONS

Vacuum pumps

type

ROTAVAC valve control

ROTAVAC valve tec

max. suction power
( 50 / 60 Hz)

m

3

/h

1,7

0,75 / 0,9

final pressure (absolute)

mbar

9

12

final pressure (absolute) with
gas-ballast

mbar

20

20

max. adm. counterpressure at
output end (absolute)

bar

1,1

adm. ambient temperature
storage
operation


°C
°C

-10 to +60

+10 to +40

admissible relative humidity for
operation (non-dewing)

%

30 to 85

dissipated power

W

180

80

idle speed
rated speed

1/min

1500

1500 / 1800

motor protection

Thermal winding protection

sound pressure level (1 m
distance)

dBA

42

42

hose shaft in / out

mm

8

size: l x w x h

mm

235 x 145 x 175

315 x 145 x 190

weight, approx.

kg

10,5

6


material in touch with media

components

ROTAVAC valve control ROTAVAC valve tec

cover, inside

PTFE, carbon fiber reinf.

PTFE

lid

ETFE

PFA

cocking disc, diaphragm

ETFE

PFA

valve

FFKM

FFKM

diaphragm

PTFE-NBR sandwich

PTFE

hoses

PTFE

PTFE

fittings / in / out

ETFE

ETFE

Advertising