E

DISASSEMBLE, FORWARD & STORE


Disassemble

Turn item OFF and disconnect mains plug.

Transportation and store

1. Store item and its components in the original box to avoid damage in transit. Close box with

adhesive tape.

2. Store item in a dry place.

Caution: avoid shocks an vibration in transit.

DISPOSAL


For disposal, please comply with your local or national regulations.
Split by metal, plastic, etc.
Packing material to be treated as described above (material split).

TROUBLESHOOTING

Vacuum pumps

defect found

reason

correction

pump won't run or
starts running and
stops right away

 electricity missing, voltage too

low ?

 overpressure in exhaust line?
 motor overload?

 check voltage, connect

electricity

 open exhaust line.
 allow motor to chill, check for

reason, turn ON again

low suction power

 centering ring not seated

correctly

 long, thin line?
 condensate in pump?

 deposits in pump ?
 diaphragm and valves broken?
 emission of gas, vapor in

process ?

 check pump check fittings. line

and recipient

 increase cross section.
 run pump with intake stud

open for a minute

 clean and check pump heads
 replace diaphragm and/or

valves

 check process parameters

pump noisy

 sucks at atmospheric

pressure ?

 cocking disc loose?
 none of the above reasons

found ?

 install hose on exhaust end

 maintain diaphragm pump
 ship pump for repair.

pump stuck or piston
rod jammed

-

ship pump for repair