Heidolph Synthesis 1 User Manual

Page 101

Advertising
background image

101

E

RZ?

ramp

values,

time

answer: RZ(x,y)=aaa,bbb,ccc,ddd,eee,fff,ggg,hhh,iii
Note: ramp number / zone need to be set before
x=ramp number,y=ramp zone

RT?

ramp

values,

temperature

answer: RT(x,y)=aaa,bbb,ccc,ddd,eee,fff,ggg,hhh,iii
Note: ramp number / zone need to be set before
x=ramp number,y=ramp zone

AP?

actual

pressure

answer: AP=xxxx

b. Entry


SZn=xxxx! with n=1...4

nominal temperature, zone n

answer: SZn=xxx

SD=xxxx!

nominal rotational speed

answer: SD=xxxx

ST=xxx!

nominal time, timer elapse

answer: ST=xxx

SB=xxx!

operational

status

answer: SB=xxx

Entry

heater

speed

ramp active

Identic

Bit

b0

b1

b2

b3

SB=000!

off off off off

SB=001! on

off

off

off

SB=002!

off on off off

SB=003!

on on off off

SB=004! off

off

on

off

SB=005!

on off on off

SB=006!

off on on off

SB=007!

on on on off

SB=008!

off off off on

SB=009! on

off

off

on

SB=010!

off on off on

SB=011!

on on off on

SB=012! off

off

on

on

SB=013!

on off on on

SB=014!

off on on on

SB=015!

on on on on


RN=x!

ramp

number

answer: RN=x

RR=x!

ramp

zone

answer: RR=x

x=1...4 zone 1...4

x=5

Identic

Advertising