E

Hit Set Ramp key once again to exit gradient setting.

Having set a non-plausible time value, you may not exit temperature
gradient setting mode with Set Ramp key. Display shows “time invalid”
and cursor marks non-plausible value that needs to be changed.
Time values from step 1 to step 9 need to be ascending.

te m p e ra tu re g ra d ie n t

0

1 0

2 0

3 0

4 0

5 0

6 0

7 0

0

5 0

1 0 0

1 5 0

2 0 0

2 5 0

3 0 0

tim e [m in ]

te

mp.

C]




10.5.2 Delete

values

For deleting a cursor-marked time value, rotate "Value" button CCW until „---„ is
displayed. Adhering temperature value automatically is set to „0“.

10.5.3 Limiting

temperatures

Within a temperature range of 20°C and 160°C, max. temperature rise / drop, in
absence of optional forced cooling, makes reference to the table below (for physical
reasons):

temperature range

max. temperature rise

max.
temperature drop

20 – 80 °C

3°C / min

1°C / min

81 – 130 °C

2 °C / min

2°C / min

131 – max °C

1°C / min

3°C / min


Deviating temperature rise values cannot be achieved within the time interval
specified; the same of course is true for temperature drop.

10.6 Activate temperature gradient

Activate temperature gradient by hitting the Ramp active key. This mode is
confirmed by the Ramp active LED. A temperature gradient set as described in

progress time, i.e. time elapsed
from start of temperature
gradient