4 reference value by gas concentration, Reference value by gas concentration – INFICON Sensistor ISH2000 HySpeed Hydrogen Leak Detector User Manual

Page 18

Advertising
background image

Hydrogen leak detector Sensistor ISH2000 HySpeed

-

Operating Instructions

18

OP-Sensistor ISH2000 HySpeed-EN-201203

EN

Note:

You will need to repeat the calibration 2-3 times to get

Calibration OK

after

changing setup or probe.

Allow at least 30 seconds between each calibration for best accuracy!

If the message “

No Gas or Unstable Signal

” is displayed repeatedly - go back to

Detection Mode and check functionality.

If

Repeat Calibration

is displayed then this means that the measured value

deviated more than 25% (can be changed between 1-100%) from the previous
calibration. Repeat the calibration procedure.

Also set the Analysis Unit to the same as the Reference Value. If you want to use
another unit you have to put a recalculation number into

Correlation Value

which

describes the relationship between the different units.

7.4 Reference value by gas concentration

The concentration of the reference gas should be near to or slightly higher

Reject

Level

the for best accuracy and always within the range of 20-50000 ppm

Set the

Reference Value

equal to the Hydrogen concentration in your reference gas.

If premixed reference gas is used, the reference value be found on the certificate of
analysis issued for the gas concentration. Also set the

Reference Unit

to the same unit

as that used to express the leak rate of the reference leak.

Example:

Reference gas concentration contains 100 PPM Hydrogen.

1- Set the

Reference Value

= 100

2- Set the

Reference Unit

= “PPM”