Hydrogen leak detector Sensistor ISH2000 HySpeed

-

Operating Instructions

38

OP-Sensistor ISH2000 HySpeed-EN-201203

EN

* entered as integer in 100’s of seconds, 1= 0.01s, 100 = 1 s, 60000 = 600s.

Transfer of parameters

Send parameters one by one; first send the specific header (for example “A”), then
wait until the RS232 transmitter unit has sent the data (approximately 20 ms), and
then send the data (for example “1.00E+01”). Data string must be ended with a
carriage return character, chr13 (dec).

Example:

“CPPM” or “C PPM”, Carriage Return (chr 13). This sets Reject Level Rate

Unit to “PPM”.

Parameters can be sent in any order you like:

If your data was received and correct, Sensistor ISH2000 HySpeed immediately
echoes (sends back) the data.

If you send a non existing header you will not receive anything.

If the data could not be converted in the Sensistor ISH2000 HySpeed you will
receive the string “CoEr”, (Conversion Error).

Note:

Remember to use capitals for the header.

APC driver installation

APC drivers are installed in the detector from a PC. All active probes need a driver to
be installed before they can be used.

For driver installation you will need the following:

APC Driver software.

File transfer cable.

PC computer with Windows 95 or later.

Timer A

D

nnn*

Timer B

E

nnn*

Timer C

F

nnn*

Timer D

G

nnn*

Purge Level

H

n.nnE+nn

Reference Value

I

n.nnE+nn

Reference Unit

J

Text string (max 12 characters)

Reference Unit (Program settings)

JUx

x=1 to 8, 1=PPM, 2=CC/S

Parameter

Header

Data