Hydrogen leak detector Sensistor ISH2000 HySpeed

-

Operating Instructions

32

OP-Sensistor ISH2000 HySpeed-EN-201203

EN

Status signal patterns

Table 8-3. Status signals for pin 14 - 18 (see “Pin configuration.” on page 31).

DET_ERROR will go high for a short time (1-5 seconds) when the detector is switched
on. It will go low when the sensor has been checked.

In normal operation, DET_ERROR = HIGH means that there is a problem with the
sensor, probe, or cable.

DET_WAIT is high when instrument is in warm-up mode after switching on power.
Instrument will also go into warm-up if there is a temporary fault in the sensor or
sensor connection.

The timing of the status signals in relation to different events is described by the
following two examples:

Example:

Input signals issued to control the APC system should have a pulse length

of at least 40 ms.

Example:

Output signals switch with a cycle time of 10 ms (0.01 s). This is the cycle

time of the APC system.

After power on

14

OUT

DET_ERROR

15

OUT

LEAK_OUT

16

OUT

DET_ON

17

OUT

DET_SIGNAL

18

OUT

DET_WAIT

19

OUT

OUT_0

20

OUT

OUT_1

21

OUT

OUT_2

22

OUT

OUT_3

23

OUT

OUT_4

24

OUT

OUT_5

25

OUT

24 VDC OUT

Pin

Type

Signal name

Signal

Function

DET_SIGNAL

Gas detected / Sensor not recovered.

DET_WAIT

High during warm-up.

DET_ON

High when detector is on.

LEAK_ALARM

Leak above Reject Level detected.

DET_ERROR

High if Probe, Sensor or Cable is broken.