The detector can print the following information, 1 date and time for power on of detector, 2 time of print – INFICON Sensistor ISH2000 HySpeed Hydrogen Leak Detector User Manual

Page 35: 3 value of all gas signals above the reject level, 4 test result: “accept” or “reject, 5 value of signal, Table 8-7. communication specifications

Advertising
background image

Hydrogen leak detector Sensistor ISH2000 HySpeed

-

Operating Instructions


OP-Sensistor ISH2000 HySpeed-EN-201203

35

EN

Printed data

The detector can print the following information:
1

Date and Time for Power on of detector.

2

Time of print.

3

Value of all gas signals above the Reject Level.

4

Test result: “Accept” or “Reject”.

5

Value of signal.

6

Result of calibration: “OK” or “Calibration Not saved”, Date and Time,
Parameter settings.

Printing of the current value can also be requested by an RS232 command (See
“Connector pin configuration” on page 33) or ordered manually by pressing

PRINT

.

Table 8-6. Probe type determines information printed.

Analysis data output

The

Analysis Data Output

option is intended for transferring test results to a

supervising computer system such as, for example, a PLC system.

Table 8-7. Communication specifications.

The data format for Analysis data output consists of nine ASCII characters. Seven
characters show the value in engineering format (See “Engineering format” on
page 19), one character shows the result of the test, and one character shows line
feed (LF).

Table 8-8. The character indicating the result of the test is one of the following.

Example:

2.5E-02R (LF)

This example is a line feed (LF), R means that the test was above the Reject Level
limit, and the value was 2.5E-02.

Probe type

Data printed

Hand Probe P50 / COMBOX with H65

1, 2, 3, 4, 6

Setting

Value

Data rate

9600 baud

Data bits

8

Stop bits

1

Parity

None

Flow control

None

Character

Result of the test

A

Accept. Previous test was below Reject Level limit.

R

Reject. Previous test was above Reject Level limit.

P

Rejected by Purging. Previous test was above purge limit
(and Reject level limit).

C

Calibration approved. Previous cycle was calibration.
Calibration was approved.

F

Calibration failed. Previous cycle was calibration.

E

Test interrupted by “Error” that occurred during cycle
(probe or sensor error etc.).

Advertising