8.15 Sensistor ISH2000 HySpeed installation

The instrument can be installed in the operator’s panel or any other flat surface of
your leak tester. Mounting brackets and panel rubber seal are delivered with the
detector. Refer to Figure 8-6.

Fig 8-6. Panel cut out dimensions.

Measurements:

Panel cut out: 262 x 127 mm (5 x 10.3 inches).

Maximum panel thickness: 8 mm (0.3 inch).

Allow an extra 20 mm (0.8 inch) on left and right side for mounting brackets.

Depth of open space in panel is at least 15 cm (6 inches) for Sensistor ISH2000
HySpeed to fit.

Installation process

Table 8-13. Install the detector as follows

CAUTION!

The detector should preferably be mounted on a
vertical surface. Tilting more than 30 degrees is not
recommended. Tilting more makes air circulation
poor resulting in increased temperature inside
detector. This will reduce contrast of display and
lifetime of lamp and electronic circuits.

NOTICE!

After installation verify that ambient
temperature is below 50ºC.

Sensistor ISH2000

Free space horisontally

262 mm (10.3 inches)

127 mm (5 inches)

Free space vertically

Step

Action

1.

Cut hole in panel according to figure above, and
remove burrs.

2.

Check that rubber o-ring is in correct position in
the grove around the edge of the detector.

3.

Put the detector in place in the panel hole.