Change Test Mode

Select the main menu Change Test Mode
as described in ”Change Test Mode” on
p. 17.

1. Press Enter.

2. Select:

Analysis Mode,
Detection Mode or
Combined Mode
by pressing < and > until you reach
the wanted mode.

3. Press Enter.

NOTE:

To change quickly from De-

tection Mode to Analysis Mode
or vice versa, press the right-hand
button three times in succession.

1

2

In Detection Mode, the signal is displayed in the form of a bar. The length of
the bar varies with the gas concentration.

In Analysis Mode the measured value is displayed in figures, ”To Quantify
Leaks” on p. 33. The default unit is in PPM but it is possible to choose other
units, see ”Leak Rate Unit and Calibration Coefficient” on p. 35.

In Combined Mode you can see the signal displayed as a bar and the measured
value in figures at the same time.

Change Test Mode

Esc < > Enter

Change Test Mode

Analysis Mode

Analysis Mode

0.5

PPM

– Volume + Menu

Esc < > Enter

3

Explanations