Calibration” on p. 20, Calibration, Explanations 5 – INFICON Extrima Ex-certified Hydrogen Leak Detector User Manual

Page 20: Calibration time, Calibration coefficient, Password protected calibration

Advertising
background image

20

EN

INFICON - Operating Instructions EXTRIMA

Calibration

Esc < > Enter

1 3

Calibrated

2012-08-25 11:04:59

Last successful calibration

Calibration Time

The number of seconds that measurement is in progress when calibrating in the

Analysis Mode. The default value is 8 seconds but values from 5 to 30 seconds

can be used.

Calibration Coefficient

Calibration parameter. The relation between the detector signal and the displayed
number is set by the Calibration Coefficient. See ”Leak Rate Unit and Calibration

Coefficient” on p. 35.

Password Protected Calibration

The calibration function can be protected using the password so that calibrating

cannot be done by unauthorised users. Factory default is no password. For more

information see page 29 or 34.

Calibration

Select the Calibration menu as described

in ”7. Menu system” on p. 17.

1. Press Enter.

2. Select:

Calibrate

Calibration Coefficient

Calibration Time or

Password Protected Calibration

by pressing < and > until you reach

the wanted mode.

3. Press Enter.

4. To change a alfanumerical value, e.g.

Calibration Coefficient, use + and .

Use > to move to next character.

5. Save the changed value by pressing

> to the far right of the set value.

Save will then appear on the bottom

of the screen.

6. Press Save to save the set value. The

setting scale will flash to confirm
the setting.

NOTE:

If Enter Password is dis-

played, this means that the set-

ting function is protected by a

password, see ”Password” on p.

34.

Calibration

Calibration Coefficient

< + – >

4

1.00E+01

Calibration

Calibration Coefficient

Undo

Save

2.00E+01

6

2

NOTE:

Press Undo to delete the

setting and revert to the previous
value.

Explanations

5

Advertising