CAUTION!

• Do not open detector! Service of this equipment may only be carried
out by service organisations authorised therefore by INFICON, Sweden.

• If the detector gets outer damage it must be controlled and repaired
by service organisation authorised by INFICON.

• Replacement of Hand Probe and Probe Cable may be carried out by
the user.

• Do not expose the probe to a hydrogen concentration higher than
0.1 % when the instrument is not switched on, this might damage or
destroy the probe sensor.

• When the instrument is switched on the sensor withstands temporary
exposure to hydrogen concentration up to 100%. Avoid long exposures
to high concentrations.

!