INFICON - Operating Instructions EXTRIMA

Certificate:

1981011

Master Contract:

241576

Project:

2531732

Date Issued:

July 17, 2012

DQD 507 Rev. 2012-05-22

Page: 2

SPECIAL CONDITIONS FOR SAFE USE “X”

Battery Charger must be CSA Certified (or equivalent), with a maximum charging voltage of 12.6 V and a

maximum charging current of 770 mA.

Notes:

APPLICABLE REQUIREMENTS

CAN/CSA-C22.2 No. 0-M91

General Requirements – Canadian Electrical Code, Part

II

CAN/CSA-C22.2 No. 60079-0:07

Electrical apparatus for explosive gas atmospheres -

Part 0: General Requirements

CAN/CSA-E60079-11:02

Electrical apparatus for explosive gas atmospheres -

Part 11: Intrinsic Safety "i"

CAN/CSA-C22.2 No. 60529:05

Degrees of protection provided by enclosures (IP Code)

ANSI/UL 60079-0:09

Electrical Apparatus for Explosive Gas Atmospheres -

Part 0: General Requirements

ANSI/UL 60079-11:09

Electrical apparatus for Explosive Gas Atmospheres -

Part 11: Intrinsic Safety “i”

ANSI/IEC 60529:2004

Degrees of Protection Provided by Enclosures (IP

Code)

MARKINGS

For the Extrima Detector and CX21 cable the following markings are provided on a min 0.02 in thick metal

nameplate, secured to the enclosure with adhesive (Refer to Drawing 1421). For the Probe the following

markings are produced by a laser-engraving method printed directly on the side of the aluminum handle:

Extrima Detector

- CSA Monogram with C US Indicator;

- Company name;

- Model number;

- Serial number or Date Code (appears on a separate nameplate);

- Certificate reference (“CSA 2007 1981011 X”)