Technical specification – INFICON Extrima Ex-certified Hydrogen Leak Detector User Manual

Page 41

Advertising
background image

41

EN

INFICON - Operating Instructions EXTRIMA

13. Technical Specification

Power supply

AC Mains Voltage

100 — 240 V 50/60 Hz

Environment

Working temperature

-20°C — +50°C (-4F - +122F)

Start up temperature

> 0°C (>32F)

Humidity

95% RH (non-condensing)

Storage temperature

0°C — +60°C (32F - 140F)

Chemical

Jet-fuel and most common petroleum
vapours

IP-Class

IP67, 30 min @1 m (IEC529)

Dimension

Net Weight

4 kg (8.8 lbs)

Overall Dimensions

H x W x D 128 mm x 240 mm x 167 mm (5” x 9.5” x
6.6”)

Application

Europe

Zone 0, 1 and 2 (mines and dust excluded) Explosion
groups IIA, IIB, IIC,
All T1, T2, T3 gases including Hydrogen and Jet-fuels

US, Canada

Zone 0, 1 and 2 (mines and dust excluded)
Class 1, Div 1, Groups A, B, C, D
All T1, T2, T3 gases including Hydrogen and Jet-fuels

Sensitivity

Range in H2 Analysis Mode

0.5 ppm — 0.2% H2

Detection Mode

1 x 10-7 cc/s (when using 5% H2 tracer gas)

Repeatability

Typical ±10% of reading + 0.3 PPM

Linearity in H2 Analysis Mode (within 0.1
— 10 x calibration
point)

Typical ±15% (within 0.5 — 100 ppm)

Battery Capacity

Operating time

7h (3h at -20°C) (-4F)

Charging time

7-8 h, flat to fully charged. Approx. 1h to
1h operating time Charge above 0C (32F)

Advertising