Response factors for gases and vapors, To use the response factors, Preparing for field operation – INFICON DataFID Portable Flame Ionization Detector for Leak Detection and Repair (LDAR) User Manual

Page 36

Advertising
background image

074-578-P1A

36

Response Factors for Gases and Vapors

To use the response factors

1. Press the CAL key and enter the response factor for the specific compound.

2. Calibrate the DataFID with zero air and 500 ppm methane as described in the section on
calibration

3. Expose the DataFID to the sample. The displayed reading is the approximate concentration of
the specific compound.

The response factors on the INFICON web site serve as a guide to concentrations measured by the
DataFID.

NOTE: It does not matter which Cal Memory is selected or which response factor is entered, DataFID

’s

response is not specific to any one compound. The displayed reading represents the total concentration
of all ionizable compounds in the sample.

Preparing for Field Operation

Field Check List

The following items should be carried into the field to reduce or eliminate instrument down time. If
you will be in the field for a single 8-10 hour day, you should include the following accessories:

o

Table 3. Check List for Field Operation

Calibration kit(s) (Part No. MX396011)

Tank(s) of calibration gas (Part No. MX396028)

Spare gas bag for zero air (Part No. MX396017)

Gas bag adapter for zero air (Part No. MX396010)

Supply of commercial zero air

Charcoal filters (Part No. MX396022, pkg. 10)

Spare inlet filters (Part No. MX396015, pkg. 25)

DataFID Instrument Manual (Part No.074-578-P1)

Advertising
This manual is related to the following products: