INFICON DataFID Portable Flame Ionization Detector for Leak Detection and Repair (LDAR) User Manual

Page 71

Advertising
background image

074-578-P1A

71

span error

span completed

Should a fault condition arise such as

“Concentration for Span Too Low” the user will be

prompted to Retry the Span point for calibration again. Users should also look to Chapter 7
Trouble Shooting to further investigate causes for fault conditions

9. When the DataFID has completed the calibration cycle the user will be prompted to select
Done.

10. The DataFID is now calibrated and ready for use. Remove the span gas bag or regulator
from the inlet.

NOTE: Calibration does not bring a user into the Run Mode automatically. After pressing the
‘done’ soft key, the user is returned to the screen that was displayed before performing the
calibration.

Example: If the user is in the User Mode screen while editing an Interval timing and the unit
flames out, simply press the 2 Cal keys, invoke the calibration cycle and, upon completion of
calibration, the unit will revert to the User Mode edit Interval timing screen again.

If the user wishes to go into the Run mode screen at any time,

select ‘meas’ and the Run Mode

screen will appear.


Concentration for Span

too low!

cancel

retry

C

alibration completed

Press ‘done’ key to exit

calibration mode

done