INFICON DataFID Portable Flame Ionization Detector for Leak Detection and Repair (LDAR) User Manual

Page 9

Advertising
background image

074-578-P1A

9

WARNING

If there is any indication of physical damage to the DataFID after it has been
dropped, the unit must be checked out by a service technician to ensure
instrument integrity.

.

THE DataFID IS NOT INTENDED TO DETECT COMBUSTIBLE LEVELS OF GASES.

THE DataFID IS CLASSIFIED FOR USE IN ATMOSPHERES CONTAINING COMBUSTIBLE LEVELS
OF GASES.


WARNING

Do not use any other accessories with the DataFID in a hazardous location

Substitution of components may affect the safety rating.


CAUTION

To reduce the risk of fire or injury to persons, read and follow these instructions:

1. All calibration, maintenance and servicing of this device, including battery charging, must be
performed in a safe area away from hazardous locations. Disconnect all power before servicing.

2. There are no operator replaceable parts for the DataFID except the exhaust frit, and sample
inlet filter.

3. There are no operator serviceable parts inside the DataFID.

WARNING

1. If the DataFID is used in a manner not specified, the protection provided by the DataFID may

be impaired.

2. Do not defeat proper polarity orientation between the DataFID and battery charger.

3. Charge the battery pack using the AC battery charger provided with or identified for use with

this product only in accordance with the instructions and limitations specified in this manual.
To charge the battery, use only the INFICON Universal Battery Charger Part No. A1201221.
When using the battery charger, do not block access to the mains outlet in use with adapter.
The AC battery charger is not to be used in a hazardous area.