zero cal

zero too high

span level

Should a fault condition arise such as

“Concentration for Zero Air too high” the user will be

prompted to Retry the Zero point for calibration again. Users should also look to Chapter 7
Trouble Shooting to further investigate causes for fault conditions

6. The DataFID then asks for the span gas concentration. Enter the known concentration or
select New to enter a different span gas concentration and then connect the gas bag adapter to
the inlet.

Selecting New will prompt the user to reset the digits for the span gas value. The left and center
keys will change the corresponding values up or down while the right key will move the place
holder over to the right. If the place holder moves too far to the right, press the right arrow key

Calibrating zero air

Concentration for Zero

air too high

cancel

retry

Span Gas concentration


500.0 ppm

cancel new

next