INFICON DataFID Portable Flame Ionization Detector for Leak Detection and Repair (LDAR) User Manual

Page 86

Advertising
background image

074-578-P1A

86

Problem: Date and time settings are not retained.

Cause: The DataFID has not been used for 3 months or more and the internal battery has been

discharged.

Action: Turn the DataFID

on and allow it to run until a “LoBatt” status appears. This will

take approximately 15 hours. Recharge the battery overnight. Repeat this
procedure for 3 or 4 days. While the DataFID is running the internal battery is
charging.

Problem: Cannot fill the hydrogen fuel cylinder.

Cause: The hydrogen supply tank has insufficient pressure. You can only fill the internal

cylinder to a pressure of less than or equal to the tank pressure.

Action: Fill the hydrogen fuel cylinder or replace the hydrogen fuel cylinder with a full
one.

Cause: There is a problem with the refill adapter.

Action: Contact the INFICON local service personnel.

WARNING

DO NOT modify or disassemble the refill adapter! If you have any
problems, the unit must be returned for repair or replacement.

Problem:

Instrument status shows “Over”.

Cause: Rapid change in signal level. The detector electronics have been

momentarily saturated.

Action:

Wait a few seconds for the status to return to “Ready”.

Cause: The detector has become saturated.

Action: Move the DataFID to a location where it can sample clean air. Sample zero air

until the reading stabilizes around 0.

Problem: Display is blank

Cause: Battery pack is critically low.

Action: Recharge the battery pack.

Cause: The battery charger is not connected to the instrument properly.

Action: Ensure the battery charger connection is attached properly to the DataFID.

Advertising
This manual is related to the following products: