2.4 Electrical Connections

WARNING

Be sure the power switch on the CI-100 controller front
panel is OFF.

Plug the power cord into the CI-100 power entry module, then into a 100-120 or
200-240 V (ac), 50/60 Hz mains outlet.

Attach the DB-25 control cable from motor drive to the connector marked Motor on
the CI-100 controller.

The I/O and RS-232/LAN connections are covered later in this chapter. They are
not needed for initial setup.

2.5 Calibration

NOTE: The display flashes E at power-up until this calibration step is completed.

WARNING

Be sure the power switch on the CI-100 controller front
panel is OFF.

CAUTION

Verify that the front panel Manual/Auto switch is set to
Manual, and the Pocket/Jog switch is set to Jog.

1

On the rear panel of the CI-100 controller set the Pocket Type and Select dip
switches to match your indexer. See the Rear Panel section of the next chapter
for a detailed description of the rear panel switches.

2

Push the CI-100 front panel Power switch to the ON position.

3

Press the CW/CCW switch to jog your indexer to Pocket 1. Press and hold the
CW/CCW switch for faster movement.

4

Using the CW/CCW switch, jog the indexer to the exact position you want to
calibrate as “home” for pocket 1.

5

Press and hold the Calibrate switch until the displays stop blinking (about three
seconds). This calibrates pocket one home.

Advertising