RQCM – RESEARCH QUARTZ CRYSTAL MICROBALANCE

GENERAL DESCRIPTION 1-1

1 GENERAL

DESCRIPTION

The RQCM is designed for many types of research applications where QCM (Quartz

Crystal Microbalance) measurement is desired. Included with each instrument is a

Windows™ based software package that allows the user to configure the RQCM, setup

multiple experiments, log data with real-time graphing and review results from previous

experiments.
The QCM portion of this system accurately measures crystal frequency and crystal

resistance for up to three crystals simultaneously. The software uses this data to derive

various physical parameters of the deposited film and/or the liquid or gas environment at

the surface of the crystal. The heart of the system is a high performance phase lock

oscillator (PLO) circuit that provides superior measurement stability over a wide

frequency range (3.8 to 6.06 MHz, or 5.1 to 10 MHz). The circuit incorporates adjustable

crystal capacitance cancellation reducing error caused by the parasitic capacitance of the

crystal, cable and fixture. Capacitance cancellation is essential for accurate measurements

of lossy (soft) films.
Data collection from external sources is accomplished with an optional Data Acquisition

Card, which provides three temperature inputs (RTD, Thermocouple and Thermistor) as

well as five scalable analog inputs. For example, you can use the analog inputs to acquire

potential and current data from a potentiostat during a cyclic voltammogram. The

RQCM allows you to combine this data with the mass data of the QCM to create a graph

of mass and current versus potential.
Control of external instruments and peripheral devices is accomplished with an optional

input/output card. Each remote I/O card provides eight remote inputs and eight relay

outputs. The functions of the inputs and outputs are defined in the RQCM’s software

with some typical uses including the control of pumps, heaters, valves, instrument

initiation, etc.

1.1 FEATURES

1.1.1

VERY WIDE FREQUENCY RANGE

The RQCM supports a wide frequency range from 3.8 to over 6 MHz. It will support

both 5 and 6 MHz crystals; and with a low limit of 3.8 MHz it will support 1.2 MHz of

frequency shift on a 5MHz crystal. A frequency range of 5.1 to over 10 MHz is also

available.

1.1.2

SUPPORT FOR VERY LOW Q, HIGHLY DAMPED, CRYSTALS

The RQCM will accurately measure crystals with resistances up to 5 KΩ. In most cases

it will maintain lock up to a resistance of 10 KΩ or more. It will support crystal

oscillation in highly viscous solutions of more than 88% glycol in water.

1.1.3

DIRECT REAL-TIME MEASUREMENTS OF CRYSTAL FREQUENCY, MASS,
AND RESISTANCE

The RQCM accurately measures crystal frequency, mass, and resistance. The

software uses this data to derive various physical parameters of the deposited film

or media at the surface of the crystal.