RQCM – RESEARCH QUARTZ CRYSTAL MICROBALANCE

OPERATION

3-6

the SMB connector labeled Crystal but don’t install a crystal.
If the Sweep LED is flashing, press and hold the Reset button and then turn the fine trimmer

clockwise until it just stops flashing. Go back and forth a few times to get a feel for the point

where the Sweep LED just stops flashing. Release the Reset button and the Sweep LED should

begin to flash again.
Install a crystal. The PLO should lock. Even so, press and hold the Reset button and again adjust

the fine trimmer to the point where the flashing just stops. The capacitance cancellation

adjustment is now perfect. Remember to check this adjustment whenever the crystal’s

environment changes.
If you could not find the proper zero capacitance point using the fine trimmer alone, then the

coarse rotary switch needs to be adjusted. Follow the instructions below to set the coarse rotary

switch.
First adjust the fine trimmer so that it is 50% meshed and the rotor plates are below the shaft.

You can see these plates through the oversize adjustment hole. See Figure 3. Next connect a

cable and crystal holder, and the crystal, if you haven’t already done so.

Figure 3 Capacitance Adjustments

Set the course rotary switch to its minimum. Depress the Reset button, and observe the yellow,

Sweep, LED. It should be flashing very fast, indicating the capacitance is grossly under

compensated. Now rotate the course switch clockwise, one step at a time. At each stop observe

the yellow, Sweep, LED, at some point it will cease to flash. Back off one stop so the flashing

begins again. The course adjustment is complete.
Slowly adjust the fine trimmer clockwise (increasing capacitance) until the flashing of the Sweep

LED just stops. The capacitance compensation adjustment is now complete. Release the Reset

button and assuming the crystal is not dead or out of range, the RQCM will lock on it.

3.5.1

AS A GENERAL RULE

The Reset switch must be depressed during the adjustment for capacitance cancellation.

Capacitance cancellation is essential for accurate measurements of liquids and lossy (soft)