RQCM – RESEARCH QUARTZ CRYSTAL MICROBALANCE

OPERATION

3-2

3.2 UNDERSTANDING

AND SETTING UP A CRYSTAL MEASUREMENT

CHANNEL

The RQCM can have up to three independent crystal measurement channels. Each channel has a

crystal input, three status LED’s, fine and course capacitance adjustments, a reset switch and a

crystal face connection. Refer to Figure 2.

3.3 FRONT PANEL DESCRIPTION

3.3.1 LOCK

INDICATOR

The green, Lock LED is on when the oscillator is locked on a crystal’s resonant frequency.

3.3.2 UNLOCK

INDICATOR

The Red, Unlock LED will be on whenever the oscillator is not locked on a crystal.

3.3.3 SWEEP

INDICATOR

The Yellow, Sweep LED flashes each time the frequency ramp is reset to its low starting point.

In normal operation, the sweep light will only flash while adjusting the capacitance

compensation. The Sweep LED will not light when locked on a crystal.

3.3.4 RESET

SWITCH

The Reset switch should be pressed while adjusting the capacitance compensation. This switch

forces the VCO to its lowest frequency independently of the Integrator output. The Reset switch

also forces the Lock LED on, thus turning off the quadrature current injection which is required

for proper capacitance cancellation adjustment. The equivalent of about 1.5 pfd of capacitance is

added as quadrature current to insure that the VCO ramps up in frequency when not locked onto a

crystal. The quadrature current is turned off as soon as a lock is detected.

3.3.5 CRYSTAL

CONNECTOR

The SMB connector labeled Crystal provides connections to the crystal. When used with a INFICON

crystal holder, the center pin connects to the crystal’s rear electrode, and the connector

housing connects to the crystal’s front electrode.

3.3.6

CRYSTAL FACE CONNECTOR

The Crystal Face connector provides a direct connection to the crystal’s front electrode. Note that

the mating connector (2.5 mm male plug) provided with each Crystal Measurement Channel must

be used for proper connection. When the mating connector is inserted, the crystal face electrode

is galvanically (transformer) isolated from earth ground allowing a potential to be applied. Use

this connection to connect the crystal face electrode to the working electrode of a potentiostat or

galvanostat for electrochemical experiments. When this connection is not in use, the crystal face

electrode is grounded to minimize effects of capacitance that may couple to crystal face electrode

in non-electrochemical experiments.