Thermocouple input, Rtd input, Thermistor input – INFICON RQCM - Quartz Crystal Microbalance Research System User Manual
Page 18: I/o card (optional), Communications, Front panel indicators, Power requirements, Physical, Thermocouple input -4, Rtd input -4

RQCM – RESEARCH QUARTZ CRYSTAL MICROBALANCE
GENERAL DESCRIPTION
1-4
1.2.2.2 Thermocouple
Input
Type:
Type “T” thermocouple
Temperature range:
0 to 371°C
Accuracy:
±
2°C + sensor error
1.2.2.3 RTD
Input
Type:
100 Ω Thin film platinum
Temperature range:
0 to 600°C
Accuracy:
±
4°C + sensor error
1.2.2.4 Thermistor
Input
Type:
100 KΩ
Temperature range:
0 to 150°C
Accuracy:
±
0.5°C + sensor error
1.2.3
I/O CARD (OPTIONAL)
Number of Discrete Inputs:
8, ground true 4.7KΩ pulled up to 5V
Number of Discrete Outputs:
8, SPST relays, 120VA, 2 A max.
1.2.4 COMMUNICATIONS
RS-232 serial port standard.
RS-485 serial port optional.
IEEE-488 optional.
1.2.5
FRONT PANEL INDICATORS
Communication Status LED’s
System Ready LED
1.2.6 POWER
REQUIREMENTS
100, 200, 220, 240 VAC @ 50/60Hz, 25 W
1.2.7 PHYSICAL
Size:
4” H (including feet) x 13” W x 9 ¾” D
Weight: 7 lbs.
Shipping Weight: 10 lbs.